Operation
Ephus
- Ephus should be configured to save the xsg file. This can be loaded as a mat file that contains the state of Ephus during the acquisition. It contains a structure called header which contains the pulse info and gain along with the actual stimulus trace
Configuration hotswitches
- Save configuration settings for program windows regarding recording and stimulation settings
-
See:
- Programs → hotswitch
- Programs → hotswitch: hs_config
-
Located in
C:/Rig/Ephus 2013b/hotswitches
Looping
- Programs → LoopGui
- Set number of iterations
- Set interval time between iterations
- Ensure stimulator and acquirer are set to external and use PFI9 for triggering
ScanImage
- 145×145 μm images at 256×256 pixel resolution at an effective frame rate of 5 Hz
Maintaining cells in field-of-view
- Use screen annotation software (such as gInk) to annotate cells so that they may be kept in focus in the field-of-view throughout the experiment.
Looping (cycles)
Cycles allow multiple iterations with defined delays between acquisitions. Cycles are saved in: C:\Rig\Scanimage5-3\cycles
Setup procedure:
- View → Cycle Controls to open cycle configuration (CYCLE-MODE CONTROLS)
- Check Cycle mode on
- Load existing cycle or save new cycle configuration
Configuring iterations:
Multiple iterations can exist for each cycle, with each iteration preceded by a defined Iteration Delay.
Example: 1 cycle with 4 iterations, each with 120s delay
- Click >> for cycle iteration data
- Cycles done: displays
X of 1 - Add 4 iterations
- Enter
131in Iteration delay field and press Enter- Calculation: Frame collection time (~11s for 50 frames at 5 Hz) + desired delay (120s) = 131s total
- Check Restore original cfg
- Check Auto-reset iteration/count
- Click CYCLE to begin cycle
Note: There is a timing delay of ~97-99 ms (occasionally up to 700 ms) between the ScanImage acquisition start message (e.g., "Starting '1 x zoom 5 Hz' at 6/13/2018 12:33:16.136") and the acquisitionStarted timestamp in the digtrig_stim user function.